Tuesday, October 14, 2008
***NOTE: This Blog has moved. Click here for the updated blog.*** Pharma Cleaning Validation – Swab Limit Calculations Question:How do you calculate the swab limit, once it has been extracted in solution? Answer: In general the calculations go as follows: swab limit (ug residue/swab) = acceptance criteria (ug residue/cm2) X swab area (cm2) X swab recovery (%) acceptance criteria = ug residue/cm2 of a given residue swab area = cm2 of area swabbed swab recovery = the percent actually detected compared to a spiked surface area of known amount of residue (amount detected/amount swabbed X 100) To convert swab limits to test result limits by converting ug residue/swab to ug residue/mL based on the extraction volume that is used to extract the swab for analysis. The general equation is: test result limit (ug residue/mL) = swab limit (ug residue/swab) / mL of extraction solution. If you need help calculating swab limits for pharmaceutical cleaning validation, contact the experts at cleaning@alconox.com. (Pharma Cleaning Validation – Swab Limit Calculations) |
Previous Posts
|
![]() |
|
| Home | Get Sample | Ask Alconox | Locate a Dealer | Literature By Mail | Site Map | About Us | Book & Sample Offer | | Customer Recommendations | Customer Purchasing | Healthcare | Pharmaceutical | Biotechnology | Medical Device | Laboratory | Environmental | | Food & Dairy | Electronic | Metalworking & Precision Mfg | Nuclear | Optical | Cosmetic | Filter | Tattoo | For Our Dealers | Alconox Site Index | Alconox, Liqui-Nox, Terg-A-Zyme, Alcojet, Alcotabs, Det-O-Jet, Detergent 8, Citranox, Luminox, Citrajet, Solujet and Tergajet are ® registered trademarks of Alconox, Inc. |
|
|